HASSAN, Ali M. and ELREEFY, SOHAIR A. and ELSALAMOUNY, AHMED R. (2011) NICKEL ION SELECTIVE POLY VINYL CHLORIDE MEMBRANE ELECTRODE BASED ON LIX 54. Al-Azhar Bulletin of Science, 22 (2). pp. 37-44. ISSN 1110-2535
ABSB_Volume 22_Issue Issue 2-A_Pages 37-44.pdf - Published Version
Download (149kB)
Abstract
PVC based ion selective electrode, which was selective and sensitive to Ni (II) ions, was developed using 1- phenyldecane - 1,3 - diones (LIX 54) as a carrier. The electrode exhibited a good potentiometric response for Ni (II) over a wide concentration range (1.0×10-5 – 1.0×10-1 M) with a slope 29.1 + 1.6 mV/decade. The proposed electrode has a response time about 25 s. The electrode is suitable for use in aqueous solutions in a wide pH range of 8.0 to 9.5. The proposed electrode was used in two different applications such as the determination of nickel concentration in nickel-cadmium battery and as an indicator electrode for potentiometric titration of Ni2+ion with EDTA.
Item Type: | Article |
---|---|
Subjects: | Impact Archive > Medical Science |
Depositing User: | Managing Editor |
Date Deposited: | 17 Jul 2023 05:16 |
Last Modified: | 10 Oct 2023 05:19 |
URI: | http://research.sdpublishers.net/id/eprint/2698 |